Basic Specifications
Measured Parameters:
C, L, R, ESR, Rs, Rp
Measuring Frequencies:
100Hz, 1 kHz, 10kHz
Measurement rate:
1 time per second, default
DC Voltage:
0 to 800 mV
(Up to 8V with optional slide switch manual setting
Resistance:
0 to 9M
Capacitance:
10 pF to 900 mF
Inductance:
1uH to 999mH
Battery life:
minimum
120 hrs (Air-Zinc batteries)
Detailed Accuracy Specifications
Accuracy is specified at 18°C to 28°C (64°F to 82°F), with relative humidity to 90%.
Resistance
Range:
0.1Ohm - 5MOhm
Accuracy:
1% in range 0.1R - 5M
Resolution:
0.01R in range 0-10R
Test Frequency:
1 kHz
Capacitance:
Range:
10pF - 499uF
Accuracy:
5% in range 20pF - 1000uF
Resolution:
0.5pF in range 1pF- 100pF
Test Frequency:
1 kHz C > 1000pF
10 kHz C < 1000pF
100Hz C > 1uF
Inductance:
Range:
1uH - 1H
Accuracy:
10% in range 1uH - 10uH
5%
in range 10uH - 100uH
5%
in range 100uH - 1H
Resolution:
0.5uH in range 1uH – 100uH
Test Frequency:
10 kHz L < 1uH
1 kHz L > 1uH
100Hz L > 1mH
Feature Summary
Dual Graphics Displays
Analog Bar Graph
Fully automatic measurement of Inductance, Capacitance and Resistance
Auto range:
Meter automatically selects best range
Continuity / Open test:
Beeper sounds for resistance readings below threshold,
or to indicate a momentary open circuit
Trace Graph: