Once the MCA has been added, then you must configure the Ethernet and memory
size of the AIM module through the Devices/MCA menu (Figure 6). The 9660 module
is thought of as three devices by the Genie-PC / Genie-2000 software: DSP Gain, DSP
Filter, and Stabilizer. The ICB address of the 9660 must be defined via the Devices
menu for any one of the aforementioned devices.
All of the programmable settings are supplied with initial default settings. Note that
some parameters can be only programmed via the MID Editor (configuration only pa-
rameters), some can be only programmed via the Adjust screen (adjust only parame-
ters), some can be programmed in both places. In any event, if the setting is adjusted,
then the new value is remembered as part of the detector input definition and used in
subsequent operations involving the detector input. To simplify setup, the various pa -
rameters have been grouped by function under the device names mentioned earlier.
MID Editor
The following section describes those parameters for the 9660 that can be accessed
from the MID Editor via the Settings menu.
Chapter 3 - Genie™ 2000 User Interface and Controls
8
Model 9660-9660A ICB Digital Signal Processor
Figure 6 Device Menu for AIM with DSP Input