Pole/Zero:
2048
Auto P/Z:
Available only on the 9660A. If equipped, do
not initiate Auto P/Z at this time.
Note
Please see “Rise Time and Flat Top Adjustments” on page 63 for additional
information on setting the Rise Time and Flat Top settings and their relation-
ship to traditional Gaussian shaping times.
8c. Stabilizer Device Adjust Screen
*Gain Mode:
OFF
*Zero Mode:
OFF
Detector Matching
The following section discusses how to adjust and optimize the Pole/Zero (can be set
automatically for Model 9660A) and use the Auto BDC function. Pole/Zero compen-
sation is extremely critical for achieving good performance at high count rates when
using detectors with RC preamps. It is equally important for good overload recovery
due to high energy and cosmic events. The Pole/Zero adjustment range accommodates
RC preamp fall times of 40
µ
s to 1.7 ms. When a reset type of preamp is used, P/Z
compensation is not required and must be set to infinity represented by a value of “0.”
If using a detector with a Reset preamp such as the Canberra 2101 or 2008, change the
preamp type from RC to TRP. With TRP selected, the Pole/Zero compensation is au-
tomatically set to a value of “0" representing infinity or no compensation. If the RC
mode is selected the P/Z SETTING must be set to “0.” Please refer to “Operation with
Reset Preamps and Selecting Inhibit Mode” on page 72 for additional information
when using Reset preamps.
Note
Once the Pole/Zero is optimized for the intended detector, the digital filter pa-
rameters (Rise Time and Flat Top) can be changed as required without the
need to make further Pole/Zero adjustments. However, the Pole/Zero compen-
sation
must
be readjusted if the detector is changed.
The Auto BDC function sets the trapezoid flat top to the correct length insuring that all
the detector charge is collected for accurate energy analysis.
Automatic Pole/Zero Matching (9660A only)
The Automatic Pole/Zero option is installed only on the 9660A. To verify the model
number please check the serial number tag located on the NIM power connector hood.
Chapter 6 - Setup and Operation
38
Model 9660-9660A ICB Digital Signal Processor