Pole/Zero Matching and “Wrap Around”
When the monitor output is processing large signals, the output may “wrap around” as
the signal reaches the maximum level. Often a clamp box such as the LB1502 can still
be used to make the pole zero adjustment (see Figure 31, where the gain has been ad-
justed to produce “wrap around”).
Appendix B - Performance Adjustments
68
Model 9660-9660A ICB Digital Signal Processor
Figure 29 Undercompensated Pole/Zero
Figure 30 Overcompensated Pole/Zero
Scope:
Vertical: 50 mV/div
Horizontal: 10
µ
s/div
Scope:
Vertical: 50 mV/div
Horizontal: 10
µ
s/div