7 Faulty Handling
7.1
Fault handling guide
The following table lists the some of the possible failure and the user can take measures:
The fault
phenomenon
The possible reasons
Take measur es
Can't boot up
Power
cord
is
not
connected
Check whether the power cord connection is
reliable
Leakage circuit breaker is
not closed
Closing the leakage circuit breakers ,
reboot . If leakage breaker disconnect
again , contact an authorized service
personnel
Fuse burn out
Replace the fuse
Prompt
abnormal
Control panel fault
Change control board, please contact the
authorized maintenance personnel
Touch
screen
no response
Touch screen failure
Please contact the authorized maintenance
personnel
Control system failure
Close the key switch to boot, such as fault
phenomenon, please contact the authorized
maintenance personnel
LCD
screen
recalibration
Contact
the
authorized
maintenance
personnel
Not pre - boot
Laser
generator
fail u re
Contact
the
authorized
maintenance
personnel
Power Failure
Replace the power supply, please contact
the authorized maintenance personnel
Control panel failu re
Change control board, please contact the
authorized maintenance personnel
Wavelength
switching
abnormal
Switching motor failure
Replace the motor, please contact the
authorized maintenance personnel
Control panel failu re
Change control board, please contact the
authorized maintenance personnel
KTP crystal loss or
damage
Maintenance KTP, please contact the
authorized maintenance personnel
Garbled ,
missing li nes Control panel failu re
Change control board, please contact the
authorized maintenance personnel
Not light , but
the
screen
display count
The pedals damage
please contact the authorized maintenance
personnel
Level set too low
Light energy level setting is too low, increase
energy levels
The life of the pump
source depletion
Replace the pump light source
Control panel failu re
Change control board, please contact the
authorized maintenance personnel