Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
b) Acne bloodshot
c) Dimmed: blood clotting
Perivascular reddish: local inflammation (two to three days to disappear)
d) Vasoconstriction, blood vessel walls collapse (color becomes shallow
gradually
Within 21 days)
e)
Tattoos, tattoo, lip liner, eyeliner: Pigments peeling, skin redness (usually
disappear
Within 2 hours), the next anti-color (deep skin reactions to the surface.)
8.6.3
The main points of guidelines
1) Eyebrow, lip liner, eyeliner ----- sensitive parts, the laser energy is low as far
as possible, the spot is too large, to avoid blasting the skin with a laser focus
2) Tattoo -- - according to tattoo color, choose suitable wavelength. Do not
happen overnight, do not repeat the same parts
3) Chloasma, age spots, using 532 nm wavelength of green light, the spotsize
as far as possible choose slants big, can not only effectively remove pigment, does
not burn the skin at the same time. Due to the capillary uptake of green light is
stronger, avoid burn skin
4) Birthmarks - identify clearly, congenital birthmark is pigmentation, or capillary
pathological changes. If the latter, the effect is not obvious
8.6.4
Operation skills
1)
For safety, always use the lower energy no upper part of the bone (for
example: the forehead, temples and cheekbones, etc.)
2) We will focus on the head close to the skin, to ensure that the spot size and
the scale corresponds to the value
3) Do not operate eye socket, the part of around the eye should be stretch to
the bones above and operate again (guarantee the location of the protective
goggles)
4) Try to avoid repeated irradiation
8.6.5 Operation process
1)
Ready to cold gel, benn plate, goggles, protective goggles, tissues, and
records
2) Clean operation area
3) Boot into the main interface of corresponding items
4) Operating parameters according to the reference parameter Settings (start
from low to high)
5) The operator and the customer should wear protective goggles
6) If there are any hair, remove
7) The skin anesthesia processing (if necessary)
8) Do the test spot (a light is perpendicular to the operation area and gently press
to ensure the most widespread contact), according to the characteristics of effective
adjustment level range and parameters were determined
9)
All bit operation, at the same time, the sensitive parts of the pain, thin parts
and skin darker parts should reduce some energy, and observe the skin reaction
30 / 32