GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 109
Note:
The below tests and values are not intended to represent treatment
parameters, but rather provide a check on proper functionality of the
handpiece and provide a reference for the user to help identify changes in the
handpiece operation.
If you use multiple delivery systems, these checks should be made utilizing
the smallest and largest distance gauge for each delivery system.
Procedure:
1.
Put on appropriate laser safety glasses
2.
Put the laser in “Standby” Caution- laser should remain in “Standby”
mode for duration of test.
3.
Select “DCD –on”.
4.
Install appropriate distance gauge from Table 6.2 below.
5.
Select “Spray Duration” shown in Table 6 -2.
6.
Place hand piece distance gauge ove r desired distance gauge spots on
the template (1301-00-8291). For 6 mm to 18 mm spot sizes, use the 8
mm to 18 mm spots on the template in the area listed for GL/GYAG
Family Spot Sizes. For the 1.5 mm and 3mm spot sizes, use the
distance gauges with the small treatment ring. Place the distance
gauge over the 5 mm spot on the template in the area listed as Vbeam
Spot Sizes.
7.
Press purge button.
8.
Remove hand piece QUICKLY from template
.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...