GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 11
Multiple handpiece delivery systems are offered with the laser system depending on
which options are purchased. The
GentleYAG
Laser System includes three delivery
systems (1.5/3 mm slider, 6/8/10 mm slider, 12/15/18 mm slider). Each delivery
system (Figure 1.2) consists of a cable assembly, replaceable distance gauge and a
handpiece assembly.
The cable assembly contains the fiberoptic, cryogen input line and valve control
wires. If the laser system is equipped with the optional DCD, the handpiece
assembly will contain the DCD spray nozzle, safety and detection electronics,
focusing lenses, and an output window to protect the lenses from dust and debris. The
spray nozzle is located near the distance gauge at the treatment end of the handpiece.
Handpiece
Delivery
System