The DCD Spray should completely fill the inner spot.
NOTE: Spray outside of the spot is acceptable as long as the inner spot
is completely filled. (This spray may be from reflected spray off the
paper). If the spot does not fill or a leak is noted, the handpiece
assembly should be replaced or contact Candela Technical Support.
Distance Gauge
Spot Size
Spray Duration
Setting
1.5 mm (small)
30 ms
3 mm (small)
30 ms
8 mm
30 ms
10 mm
30 ms
12 mm
40 ms
15 mm
50 ms
18 mm
60 ms
Note: Use the 8 mm distance gauge and 8 mm spot on the
template when testing the 6 mm spot size.
Table 6-2
Spray Settings
for Fill Test