Test #3 Cryogen Air Bubble Detection
Purpose:
To ensure that that air bubbles in the cryogen line are detected and the
associated “fault” message is displayed on the system.
Important: This procedure must be done with each GentleYAG
handpiece delivery system in your possession. The subsystem being
tested is inside of each delivery system, not in the laser system.
Procedure
1
Insure that the installed DCD canister is not empty.
Warning: Put on appropriate laser safety glasses
2.
Turn on the system and allow the “WARM UP” to
complete.
3.
With the Electrical Connector installed, disconnect the
Cryogen Connector (see Figure 3).