Page 113
8. Depress the footswitch to pulse laser continuously until a “PURGE
REQUIRED” window box appears on the user screen and lasing
ceases. This should happen in less than 50 pulses.
Results
Comments
Action
If you get
“PURGE
REQUIRED”
This particular
handpiece passed
the test
You may continue use of this
handpiece
If “PURGE
REQUIRED”
does not occur
This particular
handpiece failed
the test
Discontinue use of this
handpiece and call Candela
Tech Support
at (508) –358-
7637 ex 336 or (800) –733-
8550 ex 336
9.
Reconnect the Cryogen Connector (see Figure 3) and press reset to
use the system, or repeat the test on other GentleYAG handpiece
delivery systems.
10. Depress “purge” may be required until the line is refilled.