SITUATION
/ SYMPTOM
PROBABLE
CAUSE or
INDICATOR
SOLUTION
The system
cannot be
turned on.
The power is not
connected
properly.
Reseat the power cable
and check circuit breaker.
The laser system
circuit breaker is
in the "off"
position.
Switch the circuit breaker
to the "on" position.
The keylock
switch was not
fully engaged.
Turn the keylock switch
fully clockwise to the "
" position and release.
The external
interlock is
defeated.
Check the remote
interlock connection. If
connected to a door, make
sure the door is closed.
Laser pulses,
no cryogen is
delivered
The DCD Pre and
Post spray
settings are set to
zero “O”.
Select the DCD Pre or
Post spray and use the “up
arrow” to increase the
spray setting.
TROUBLE-
SHOOTING