GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 117
SITUATION
/ SYMPTOM
PROBABLE
CAUSE or
INDICATOR
SOLUTION
Cryogen leak.
Tubing breaks in
the delivery
system.
Remove the cryogen
canister or disconnect the
handpiece assembly from
the laser. Call Service.
Warm-up time
has exceeded
60 minutes.
The water
temperature
control circuitry
failed.
Call Service.
Ineffective
fluence
response.
System or Fiber
is degraded.
Perform a calibration
procedure per Section 5.
Call Service if problem
persists.
Replace
Canister
Message
Appears
There is
insufficient
cryogen in the
canister.
Replace the cryogen
canister with a new
canister supplied by
Candela; Depress the
Canister Count switch for
3 seconds to reset the
canister count.
TROUBLE-
SHOOTING