GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 122
9.1
DI temperature < 62degC after 60 minute warm-up
9 – WARM -UP
TIMEOUT
9.2
DCD pressure < 105psi after 60 minute warm-up (Only
when DCD enabled)
10.1 HP type changed or unrecognized while in Ready
10.4 Fiber not detected while in Ready
10 – DELIVERY
SYSTEM
FAULT
10.5
Slider button pressed while in Ready State. Cannot change
spot -size while in Ready.
12.1
On last treatment pulse, the head energy (HD) was 14%
lower than expected head energy (xHD).
12.2
On last treatment pulse, the head energy (HD) was 14%
higher than expected head energy (xHD).
12.3
The head energy (HD) of the last pulse > Max Treatment
HD Energy of 105J.
12 – ENERGY
OUT OF
RANGE
FAULT
12.4
The HD total energy is not evenly balanced between each
subpulse (+-20%)
13 – TRIGGER
SWITCH
FAULT
13
The redundant trigger switches were in two different states
for > 1 second while in ready.
14 – SIMMER
FAULT
14
The simmer did not start or dropped out while in ready.