GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 123
15.1
Transmission (Tx) is < 70% w/HD or CP energy >1.5J
(Aper < 30% w/HD>6J or CP energy>2.5)
15 TRANSMISSION
FAULT
15.2
Transmission (Tx) is > 100% w/HD or CP energy >1.5J
(Aper >50% w/HD>6J or CP energy>2.5)
18.1 HD CktCal Test Energy incorrect or chksum corrupt.
18.2 DI Factor chksum corrupt
18 CIRCUIT CAL
FAULT
18.3 DCD Factor chksum corrupt
19.1 IGBT Trigger Fault
19.2 7875uf capacitor charge status incorrect
19.3 Lasing Timer Fault
19.4 Lasing Power Fault
19– MODULATOR
FAULT
19.5 HV Dump Fault