8501-00-1766 Revision F
CONFIDENTIAL
Page 155
3.
Install the delivery system. Set the slider for 8mm, 12mm, or 15 mm.
NOTE: When pulsing into the Ophir energy meter, the meter head
must be 6” from the handpiece to prevent damage to the meter head.
4.
There are 3 typical icon screens (shown at left) that are displayed during
each of these Circuit Cal’s and they are described below.
1)
Pulse into the Ophir meter
Pulse once when the HVr and HVs voltages are within 20V of each
other.
The handpiece is directed into the Ophir meter
2)
Enter the Ophir energy using keypad
Always enter to two decimal places.
3)
Pulse into Cal Port
Note the HVr (reference or set voltage), HVs (sample or actual voltage), HD
(Laser Head energy [J]), CP (Cal Port energy [J]) are displayed to the right of
each screen.
LASER HEAD (HD) CIRCUIT CAL (0-120J, 0-20J RANGES )
The Circuit CAL basically pulses the laser into an external meter and then into
the systems Cal Port at a low and high energy to calculate a slope and offset
difference between the two. There are three steps that are completed during the
0-120J Circuit CAL. First, the energy measurement circuit pregain (PreG) is
measured at a low energy, then a low point (LoPt) energy is measured, and
finally a high point (HiPt) energy is measured to allow for calculation of the
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...