GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 156
slope and offset. The 0-20 Circuit CAL uses the pregain from the 0-120J range
so only requires the LoPt and HiPt measurements.
5.
Press the “
0-120J
” button beneath the “
HD
” column for 1 second to begin
the HD circuit calibration procedure. The system enters Ready
automatically.
6.
The screen will prompt the user (with Icons) when to pulse the laser into
the Ophir energy meter, when to enter the Ophir meter reading (using the
keypad), and when to pulse into the calport.
NOTE: Set the Ophir meter on the 30J range. After completing the LoPt,
change the Ophir meter to the 100J range.
7.
On completion, the laser “dumps” (popping sound) the stored energy and
the screen displays “SUCCESSFUL HD CAL”
8.
Press the “
0-20J
” button beneath the “
HD
” column for 1 second to begin
the HD circuit calibration procedure.
9.
The screen will prompt the user (with Icons) when to pulse the laser into
the Ophir energy meter, when to enter the Ophir meter reading (using the
keypad), and when to pulse into the calport.
NOTE: Set the Ophir meter on the 30J range for this entire CAL.
10.
On completion, the laser “dumps” (popping sound) the stored energy and
the screen displays “SUCCESSFUL HD CAL”
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...