GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 158
15.
The screen will prompt the user (with Icons) when to pulse the laser into
the Ophir energy meter, when to enter the Ophir meter reading (using the
keypad), and when to pulse into the calport.
NOTE: Set the Ophir meter on the 30J range for this entire CAL.
16.
On completion, the laser “dumps” (popping sound) the stored energy and
the screen displays “SUCCESSFUL CP CAL”
VERIFICATION OF CIRCUIT CAL FACTORS
The Circuit CAL for all internal energy circuits has now been completed. The
energy circuit factors should be verified prior to exiting the Circuit CAL
screens. The Factors screen displays all the pertinent Circuit CAL factors as well
as provides a Test button to ensure that the Circuit CAL’s were done properly.
17.
Press the Factor button.
18.
Verify the pregain is not exactly 15000, and is between 12000 and 25000.
19.
Verify that the slope (m factor) for both ranges of the HD and CP are 1 +-
10% and that the offset (b factor) for both ranges of the HD and CP are 0 +-
15%.
20.
Verify the Aper Factor is 0.700 +-20%.
21.
Press the Test button. Verify that “Pass” is displayed in all eight locations.
22.
If any of these verifications fails, the Circuit CAL needs to be repeated. If it
fails more than once, then contact Candela Customer Service for service.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...