ELECTRICAL REQUIREMENTS
Table 1-2 lists Electrical Requirements for the
GentleYAG
Laser System.
CAUTION!
IF A PLUG OR LINE CORD NEEDS TO BE CHANGED, IT MUST BE DONE
BY A QUALIFIED PERSON IN ACCORDANCE WITH THIS SECTION AND
THE ELECTRICAL CODE OF THE INSTALLATION SITE.
The
GentleYAG
Laser System is s hipped with a twelve foot (3.7 meter) power
cord terminated with a locking NEMA L6-30P plug for power connection in the
United States. The installation site requires a mating NEMA L6-30R power
receptacle located within ten feet (3 meters) of the intended laser system
location. See Table 1-2 for electrical service requirements.
For International installations, the power connections should be made with a
grounded 2-conductor plug and receptacle pair. The plug & receptacle must be rated
for the service line voltage at a minimum, and capable of handling 3680 VA (see
Table 1-2 for detailed ratings). A plug meeting these requirements must be installed
onto the laser system line cord. Alternately, the entire line cord may be replaced with
one which is terminated with the appropriate plug
.
ELECTRICAL
REQUIREMENTS
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...