Keylock Switch
This key-operated switch controls electrical power to the laser system. The
GentleYAG
Laser System can be operated only with the key provided by Candela.
The keylock switch has three positions: “
” (off), “
” (on), and “ ” (start). To
start the laser, turn the key from the “
” position to the “ ” position, then release.
The switch spring returns to the “
’’ position. The laser system starts in a few
seconds and four quick audible beeps will sound.
Laser Stop Switch
When the red Laser Stop switch is pressed, the laser system shuts down immediately.
To restart the system, turn the keylock switch to the “ ” position, and then release.
Laser Aperture
The laser aperture is located at the distal end of the handpiece (see Figure 6.3).