GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 61
Available
Pulsewidths
(ms)
Minimum
Fluence (J/cm
2
)
Maximum
Fluence (J/cm
2
)
Maximum
Repetition Rate
(Hz)
0.25
6.0
23.0
7 (note 2)
0.30
6.0
27.0
10 (note 2)
0.35
6.0
30.0
10 (note 2)
0.40
6.0
30.0
10 (note 2)
0.45
6.0
30.0
10 (note 2)
0.50
6.0
30.0
10 (note 2)
3.0
35.0
200.0
2 (note 4)
5.0
35.0
200.0
2 (note 4)
10 to 100
(10 ms
increments)
35.0
200.0
2 (note 4)
125-300
(25 ms
increments)
35.0
200.0
2 (note 4)
Notes:
1. Fluence increments are 1 J/cm
2
from 6 to 30 J/cm
2
and 5 J/cm
2
from 35 to 200
J/cm
2
.
2. Maximum Repetition Rate = 3 Hz for 6 -11 J/cm
2
, 5 Hz for 12-20 J/cm
2
, 7 Hz for
21-23 J/cm
2
, 10 Hz for 24 -30 J/cm
2
3. The repetition rate setting is a maximum setting. The actual repetition rate
achieved may be less depending on selected fluence, pulsewidth and DCD settings.
4. The maximum repetition rate is 1.5 Hz at fluences greater than 110 J/cm
2
with a 3
ms pulsewidth. Longer pulsewidths reduce the achieved repetition rate.
Table 3.3
6 mm Fluence,
Pulsewidth,
Repetition Rate
Ranges
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...