GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 64
Available Fluence Settings for the 12 mm Spot (J/cm
2
)
10
22
34
46
58
12
24
36
48
60
14
26
38
50
62
16
28
40
52
64
18
30
42
54
66
20
32
44
56
68
70
Notes:
1. Repetition rate can be set to 0 (single pulse), 0.5, 1.0, 1.5, and 2.0 for all fluences.
The repetition rate setting is a maximum setting. The actual repetition rate
achieved may be less depending on selected fluence, pulsewidth and DCD settings.
2. The available pulsewidth settings for each fluence are: 3 ms, 5 ms, 10-100 ms (in
10 ms increments), and 125 ms to 300 ms (in 25 ms increments). 100 ms
maximum for LE.
3. The maximum repetition rate is 1.5 Hz at fluences greater than 26 J/cm
2
with a 3
ms pulsewidth. Longer pulsewidths reduce the achieved repetition rate.
Table 3.6
12 mm Fluence,
Pulsewidth,
Repetition Rate
Ranges
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...