GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 67
The following operating parameters will only be
displayed when the optional
Dynamic Cooling Device is installed.
The DCD Pre Spray Adjust parameter adjusts the duration of the cryogen spray
applied to the patient before the laser pulse. The DCD Pre Spray can be turned off
(“O”) or set to a duration from 20 milliseconds to 100 milliseconds in increments of
10 milliseconds. The 1.5/3 mm delivery system can access three additional pre-spray
settings of 10, 15, and 25 ms. Also, to access pre spray times above 30 ms with this
delivery system, hold down the UP arrow on the display for at least 5 seconds.
CAUTION
DCD spray is not allowed at rep rates greater than
2 Hz.
The DCD Delay Adjust parameter adjusts the duration of the time between the DCD
cryogen spray and the laser pulse. The delay selectable range is 3, 5, 10-150
milliseconds, in increments of 10 milliseconds.
DCD Pre Spray
DCD Delay