GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 71
CAUTION!
FAILURE TO INSTALL THE APPROPRIATE SIZE CANISTER FOR YOUR LASER
OR FAILURE TO REPLAC E IT AS INSTRUCTED CAN LEAD TO ADVERSE
CLINICAL OUTCOMES INCLUDING BURNS.
SUCH ADVERSE RESULTS MAY OCCUR AS A RESULT OF THE FOLLOWING:
•
SIGNIFICANTLY REDUCED COOLING OF THE EPIDERMIS FOR A
GIVEN LASER ENERGY
•
INADEQUATE PRESSURE TO FILL A TREATMENT AREA
•
DO NOT ATTEMPT TO US E THE CANISTER BEYOND THE FIRST
OCCURRENCE OF THE “REPLACE CANISTER” MES SAGE.
•
DO NOT RESET SYSTEM PULSE COUNTERS WITHO UT REPLACING
THE CANISTER.
•
DO NOT INSTALL PARTIALLY USED CANISTERS.
•
YOUR SYSTEM HAS BEEN CONFIGURED FOR A SPECIFIC SIZE
GENTLECOOL
TM
CANISTER. ONLY INSTALL THE APPROPRIATE SIZE
CANISTER AS INDICATED BELOW:
Canister Size
Candela P/N
Laser Type
1000 gram
1600-00-0210
GentleYAG