The calibration port (Cal Port) is used to measure the laser output energy. The
handpiece must be inserted into the calibration port in order to initiate this procedure.
The distance gauge must be removed, and the handpiece cleaned and dried before the
handpiece is placed in the calibration port.
The remote interlock connector, located on the upper rear panel, may be connected to
one or more switches on the laser room door(s). When the interlock is connected, the
laser system shuts down if the laser room door(s) is opened. The switch must be
connected so that with the door closed, the switch contacts are closed. When the
door is open,
the switch contacts must open. When remote interlock is not in use, the
supplied jumper must be plugged into the interlock connector. The circuit requires a
switch with a minimum rating of 24 VDC at 250 mA (1/4 Amp).
The footswitch connector is located on the rear panel of the laser system. The
footswitch connector is a rigid tube with a smooth taper. To install the footswitch,
push the pneumatic footswitch hose over the connector (tube). Pull to remove. The
length of the footswitch tubing can be shortened, if desired, using scissors.
The on/off main switch is located on the rear panel of the laser system and must be in
the ‘1’ (on) position for the laser system to operate. Always place the on/off main
switch in the ‘0’ (off) position when the laser is not in use.
Calibration Port
Remote Interlock
Footswitch Connector
On/Off Mains Switch
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...