GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 76
1.
Cover treatment room windows with an opaque material to prevent unintended
viewing.
2.
Post laser warning signs at each entrance of the treatment room.
3.
Ensure an adequate number of protective eyewear is available. Proper eyewear
will filter light at a wavelength of 1064 nm with an O.D. of 6.3 or greater.
4.
Plug the laser into the correct electrical outlet. Ensure that the main circuit
breaker on the upper rear panel is in the “ON” position.
5.
Select and install the desired delivery system. Insure that the fiber connector at
the laser is secure.
6.
Set up the delivery system on the fiber pole as per section 1 of this manual.
7
.
Inspect and verify that the handpiece window is clean.
8.
The keylock switch has three positions: “
” (off), “
” (on), and “ ”
(start). The switch returns to the “
’’ position once the laser system starts and a
“power on” audible beep will sound. To start the laser system, turn the
keyswitch fully clockwise from the
to the start position, and then
release. Once released the keyswitch will return to the on “
” position similar
to an automobile ignition switch . There may be a delay of several seconds
before the system initializes. This is normal. The system will now enter the
warm-up state (approx. 25 min). After the warm-up is complete, the system will
enter the STANDBY state.
LASER SYSTEM
START-UP
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...