GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 78
Note: Repetition rate may be adjusted before or after calibration
Note: Before calibration a message will appear on the touchscreen
for the user to “Confirm Treatment Parameters”. After confirming,
the “check box” must be pressed prior to proceeding with the
calibration.
g.
Remove calibrated handpiece when CAL is complete.
Note: After calibration the laser will return to STANDBY mode for
safety reasons. This is to allow the user to install distance gauge.
13.
Install distance gauge and press the READY button. Aim the handpiece at a
white piece of paper and inspect the aiming laser for circular uniformity and
clarity.
Note: If the aiming laser spot is not uniform, check for distance gauge
interference. Replace bent distance gauge if necessary. If using the 1.5 mm
distance gauge, ensure that the aiming beam is at or near the middle of the
distance gauge ring. If not, remove and install the distance gauge again to
correct the problem. The distance gauge must be replaced if correct results
can not be achieved.
Important !
Do not operate laser if aiming beam is not present! This may be an
indication of a broken fiber optic. If the aiming laser is not present, replace
the delivery system. If this does not correct the problem, call for Service.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...