GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 80
NOTES
•
To return the pulse counter to zero, press the PULSE COUNT button for
3 seconds.
•
A calibration will automatically be required when:
1)
Laser is turned on;
2)
FLUENCE parameter or pulsewidth parameter changed;
3)
Delivery system changed;
4)
Slider plunger depressed;
5)
Slider position changed;
6)
Specific faults occur;
7)
In STANDBY for more than 30 minutes;
•
The user must remember to initiate a calibration after:
1)
Cleaning or replacing a window in the handpiece Slider.
2)
Replacing the delivery system.
3)
Changing the treatment spot size