CAUTION!
FAILURE TO PERFORM CALIBRATION AFTER A HANDPIECE
SLIDER WINDOW HAS BEEN CLEANED OR REPLAC ED, CAN
RESULT IN DELIVERY OF FLUENCES GREATER THAN
SPECIFIED ON THE CONTROL PANEL.
The
GentleYAG
Laser System requires that the laser be calibrated prior to each
patient treatment. During calibration, the handpiece without the Distance Gauge is
inserted into the Cal Port allowing an internal energy meter to measure the laser
output energy delivered at the handpiece. The system
adjusts itself until the desired
output is obtained. Usually 3 - 15 laser pulses are required before calibration is
complete.
•
Inspect and verify that the handpiece Slider window is clean.
•
Select the desired Delivery System. Make sure the Delivery System is
properly connected and secured.
•
Put on laser safety eyewear.
•
Select the desired Fluence parameter values via the Display Panel.
•
Remove the Distance Gauge from the handpiece.
CALIBRATE
PROCEDURE