NOTE
If the desired fluence cannot be reached, a FAULT will be displayed. If this
occurs, decrease the fluence and perform another calibration. When this
happens, the laser system is degraded and does not have sufficient energy to
calibrate properly. A dirty window on the handpiece slider may be the cause of
the problem. Clean or change the window using the instructions in Section 6
(Maintenance and Troubleshooting section). If the problem cannot be corrected,
the system may require service. Call Candela Service for more information.
If a higher fluence is desired immediately, change the delivery system to one
with a smaller spot size.