or replace the distance gauge and slider windows before repeating this test. Use of a
damaged fiber-optic delivery system is dangerous and must be avoided. If damage is
suspected discontinue use immediately.
Always cap the proximal connector of the fiber with the attached rubber cap
whenever the fiber is not installed on the laser.
Windows have been designed into distance gauges (with the exception of the 3mm
distance gauge) and slider to protect the delivery system optics. Due to the nature of
some procedures, the windows will require frequent cleaning and/or replacement to
maintain proper system performance. Windows should be maintained in accordance
with Candela procedure 8502-00-0847. Assembly pictures and procedures specific to
the
GentleYAG
delivery systems are included in this section.
WINDOWS
Washer
Window
O-ring