GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 90
CAUTION!
ONLY USE GENTLEYAG REPLACEMENT WINDOWS IN HANDPIECE
AND SLIDER OR PERMANENT DAMAGE WILL OCCUR.
THE DISTANCE GAUGE ON THE END OF THE HANDPIECE MAY
BECOME SOILED WITH NORMAL USAGE. TO ENSURE PROPER
FLUENCE DELIVERY, IT IS IMPORTANT TO INSPECT AND
CLEAN THE DISTANCE GAUGE WINDOW FREQUENTLY SO
DEBRIS IS NOT BURNED INTO THE WINDOW SURFACE.
ALWAYS PUT THE LASER SYSTEM INTO “STANDBY” OR “OFF”
WHEN CHANGING A DELIVERY SYSTEM OR DISTANCE GAUGE
WINDOW OR WHEN A HANDPIECE SLIDER WINDOW IS
CLEANED OR REPLACED.
WHEN THE HANDPIECE SLIDER WINDOW OR THE DISTANCE
GAUGE WINDOW BECOMES DIRTY OR BURNT, THE AMOUNT
OF ENERGY DELIVERED TO THE PATIENT MAY BE REDUCED.
Distance Gauge and
HP Slider Window