GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 94
3.
The Distance Gauge window is held in a groove in the front (treatment) end of
the assembly with an O-ring. A notched access in the groove allows easy
removal of the O-ring and window. Note that the non-treatment end of the
distance gauge contains a lens. This lens is not removable by the customer. If the
lens becomes dirty, use the cleaning process at the end of this procedure.
4.
With the treatment end of the distance gauge pointing upwards, remove the O-
ring with tweezers, or poke a pointed object into the notch. Gently pull the O-
ring toward the center of the window to free the O-ring from the groove.
5.
Turn the assembly upside down, allowing the window to fall out onto a clean
surface. (If needed, gently tap the side of the Distance Gauge with your finger).
6.
(Used windows only)
a.
Clean the windows in an alcohol solution. Rinse thoroughly with clean water,
and dry with a lint free tissue.
b.
Re-inspect the window and if unacceptable, discard window and replace with
a new one.
7.
Grasp the new or cleaned window by the edges and place it back into the
Distance Gauge assembly so it is resting flat on the ledge.
8.
Reinsert the O-ring into the groove. Use the tip of the tweezers or a pointed
object to gently push the O-ring fully into the groove, being careful not to touch
the window.
9.
Perform a Calibration Procedure per section 5. Then slide the distance gauge
back into the HP.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...