To clean the 1.5 mm Distance Gauge lens: (figure 6.2)
1.
Wear dustless gloves to prevent smudges or fingerprints on lens.
2.
Put the laser in STANDBY and remove the distance gauge assembly from
the handpiece.
3.
The Distance Gauge lens is on the non-treatment end of the distance gauge.
The lens can not be removed by the customer but it can be cleaned while it
is in the distance gauge.
4.
Wipe the lens ONCE with a tissue dampened with alcohol and then discard
the tissue.
5.
Use a second tissue that is dry and wipe the lens ONCE to clean off the
alcohol. Discard the tissue
6.
Repeat steps 4 and 5 (wiping only once with each tissue) until the dirt on
the lens is removed or no improvement is seen.
7.
In most cases, only the side of the lens that is exposed on the non-treatment
end of the distance gauge will become dirty. The side of the lens that is on
the treatment end of the distance gauge should remain clean since it is
protected by the window in the distance gauge. However, if this side of the
lens becomes dirty, remove the distance gauge window and then perform
steps 4 and 5 until the lens is clean.
8.
Perform a Calibration Procedure per section 5. Then slide the distance
gauge back into the HP.
Summary of Contents for GentleYAG
Page 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Page 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Page 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Page 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Page 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...