background image

Candela Corporation  

M

GentleLASE

 Aiming Beam Alignment Procedure 

8503-01-0828, Revision 01 

PROPRIETARY 

Page 3 of 4 

CAUTION !!

 

During this procedure it is possible that the laser may fire, emitting laser radiation. 
Therefore SAFETY EYEWEAR must be worn. The eyewear must have an optical 
density of at least 5.8 at 745 - 765 nm.   

The high voltages present in this system can be LETHAL.  This procedure must 
be performed only by those technicians who are familiar with the precautions 
required when working with high voltage systems, and those who have been 
trained on the MGL system and its particular hazards.   

 

WARNING!!!!! 

BEFORE WORKING NEAR THE HEAD, ALWAYS DUMP THE VOLTAGE BY 
SELECTING CHRG DISABLE FROM THE TOGGLE SCREEN, PULSING ONCE, 
THEN ENTERING STANDBY.  THERE WILL STILL BE ABOUT 150 V ON THE 
FLASHLAMP CATHODE. THIS MUST BE DISCHARGED WITH THE 
DISCHARGE STICK.  ALWAYS MEASURE THE VOLTAGE AT THE FLASHLAMP 
CATHODE BEFORE WORKING ON THE LASER RAIL. 

 

 

1. 

Enter Maintenance Mode, press the Charge Enable button on the HV CTRL 
screen to disable the charge (not highlighted).  Pulse once to discharge the laser,  
then discharge the remaining voltage using the discharge stick. 

2. 

Install fiber receptacle crosshair plug (7122-00-3363) into fiber focusing lens 
assembly. 

3. 

Go to the toggle screen and set the aiming beam intensity to 100%.  Enter 
READY to activate the Aiming beam. 

4. 

Adjust the screws on the aiming beam gimbal mount to center the beam on the 
crosshairs. 

5. 

Go to STANDBY.  Press the Charge Enable button on the HV CTRL screen to 
return the charging circuitry to normal operation. 

  