Candela Corporation
M
GentleLASE
Aiming Beam Alignment Procedure
8503-01-0828, Revision 01
PROPRIETARY
Page 3 of 4
CAUTION !!
During this procedure it is possible that the laser may fire, emitting laser radiation.
Therefore SAFETY EYEWEAR must be worn. The eyewear must have an optical
density of at least 5.8 at 745 - 765 nm.
The high voltages present in this system can be LETHAL. This procedure must
be performed only by those technicians who are familiar with the precautions
required when working with high voltage systems, and those who have been
trained on the MGL system and its particular hazards.
WARNING!!!!!
BEFORE WORKING NEAR THE HEAD, ALWAYS DUMP THE VOLTAGE BY
SELECTING CHRG DISABLE FROM THE TOGGLE SCREEN, PULSING ONCE,
THEN ENTERING STANDBY. THERE WILL STILL BE ABOUT 150 V ON THE
FLASHLAMP CATHODE. THIS MUST BE DISCHARGED WITH THE
DISCHARGE STICK. ALWAYS MEASURE THE VOLTAGE AT THE FLASHLAMP
CATHODE BEFORE WORKING ON THE LASER RAIL.
1.
Enter Maintenance Mode, press the Charge Enable button on the HV CTRL
screen to disable the charge (not highlighted). Pulse once to discharge the laser,
then discharge the remaining voltage using the discharge stick.
2.
Install fiber receptacle crosshair plug (7122-00-3363) into fiber focusing lens
assembly.
3.
Go to the toggle screen and set the aiming beam intensity to 100%. Enter
READY to activate the Aiming beam.
4.
Adjust the screws on the aiming beam gimbal mount to center the beam on the
crosshairs.
5.
Go to STANDBY. Press the Charge Enable button on the HV CTRL screen to
return the charging circuitry to normal operation.