background image

Candela Corporation  

M

GentleLASE

 Installation Procedure 

8503-01-0821, Revision A 

Candela Corporation Proprietary 

Page 1 of 8 

MGentleLASE 

Installation Procedure 

Summary of Contents for MGentleLASE

Page 1: ...Candela Corporation MGentleLASE Installation Procedure 8503 01 0821 Revision A Candela Corporation Proprietary Page 1 of 8 MGentleLASE Installation Procedure ...

Page 2: ...MGentleLASE Installation Procedure Candela Corporation Page 2 of 8 Candela Corporation Proprietary 8503 01 0821 Revision A ...

Page 3: ...tem from the crate 2 Remove the accessories located on each side of the laser system from the crate 3 Place one end of the front of the crate on the floor Butt the other end against the floor of the crate Grasp each handle and gently roll the laser out of the crate WARNING Rear wheels must be locked when removing laser from crate Failure to have rear wheels locked could result in the laser coming ...

Page 4: ...nch of top of reservoir Always use distilled water when filling the reservoir Using water that is not distilled will result in poor flashlamp performance and may result in permanent flashlamp damage 3 Replace the black cap on the reservoir 4 Plug the male end of the power cord into the approved outlet 5 The fluid lines within the laser must now be filled with water To do this turn the system on fo...

Page 5: ...ore installing into the system 2 Insert canister threaded side down into the opening on the top left side of the laser system Turn canister clockwise until hand tight A small amount of cryogen might vent out The sound of the escaping gas may be quite loud Only after the canister is fully seated will the gas stop venting out 3 Place the cylindrical canister cap over the cryogen canister Display Pan...

Page 6: ...e F O Ring 1 Remove the handpiece assembly from the accessory kit package 2 With the system turned off place the distal end of the handpiece assembly into the calport 3 Connect the handpiece control round electrical plug connector to the handpiece control receptacle on the front of the laser system 4 Connect the cryogen output connector to the cryogen output receptacle on the front of the laser sy...

Page 7: ...locations 1 With the system turned off remove the protective cap on the proximal fiber receptacle located on the front of the system 2 Remove the fiber assembly from its package 3 Remove the cap from the distal end of the fiber Slide the distal end of the fiber into the HP Slider until it clicks firmly into place 4 Insert the proximal fiber connector into the proximal fiber receptacle located on t...

Page 8: ...doors ramps and other uneven or sloping floor surfaces are encountered A severe physical shock could cause the alignment of the laser head or the optical fiber to be disturbed Furthermore if the system is allowed to get out of control when being moved personal injury or physical damage could result If it becomes necessary to relocate the GentleLASE PLUS call Candela or the distributor for details ...

Reviews:

Related manuals for MGentleLASE