Installation Procedure
8503-01-0821, Revision A
Candela Corporation Proprietary
Page 7 of 8
mating connector into the knurled connector until it stops. Release the knurled
connector.
Distance Gauge Installation
Refer to figure 2 above for distance gauge location.
Note
The distance gauge must be removed before performing
the calibration procedure.
1.
Insert the distance gauge into the distal end of the handpiece delivery system.
Fiber Assembly Installation
Refer to figure 1 for fiber assembly system connection locations.
1.
With the system turned off, remove the protective cap on the proximal fiber
receptacle located on the front of the system.
2.
Remove the fiber assembly from its package.
3.
Remove the cap from the distal end of the fiber. Slide the distal end of the fiber
into the HP Slider until it clicks firmly into place.
4.
Insert the proximal fiber connector into the proximal fiber receptacle, located on
the front of the system, until it clicks into place and stops.
5.
Refer to the Operator Manual for a description of the controls and indicators.
Footswitch Installation
Refer to figure 1 for foot/fingerswitch connection locations.
1.
Remove the footswitch from its package.
2.
Attach the tubing end to the bulkhead connector on the rear panel of the laser.