CANNON

®

 CT-1000 Constant Temperature Bath

Revision 3.1a—February, 1998; 

CANNON

®

 Instrument Company

P.O . Box 16 • State College, PA • USA

NOTE

Two screws located on either side
of the opening for the motor stirrer
serve as locating pins for the
motor support pad. Do not remove
these screws; the holes in the pad
fit loosely over their heads. The
motor line cord should point toward
the rear of the bath (offset slightly
to the right or left). The motor
stirrer should now lie flat on the top
of the bath.

3. Reattach the top heater housing , making sure that the heater, motor,

and fan cords pass through the left-hand opening (as viewed from the
rear) and that the control probe, over-temperature probe, and level
switch cords exit from the right opening (as viewed from the rear).
The back lip on the rear top cover fits into the slot on the top heater
housing. Line up the holes, insert screws, and tighten.

4. Connect all plugs and probes to the correspondingly labeled sockets

at the rear of the CT-1000 bath unit.

5. Adjust the four feet on the bottom of the bath housing to level the

bath. This should be done

 before

 filling the bath with fluid.

Inserting viscometer tubes/thermometers

The top cover of the CT-1000 contains seven apertures, 51 mm (2

)

in diameter, for the insertion of viscometer tube holders. Two additional
holes are provided for insertion of thermometers.

Inserting viscometer tubes

If necessary, remove the viscometer tube hole cover(s) from the top of
the bath and carefully place the viscometer tube(s), with the proper
holder attached, into the bath through the aperture(s) in the top cover.

NOTE

After filling the bath with fluid, adjust the height of the viscometer tube(s)
to ensure that the liquid under test and/or any timing marks on the tube
are a minimum of 6 mm (¼") below the top level of the liquid.

Thermometer immersion

Proper thermometer immersion is critical for viscosity measurements.
Even a calibrated thermometer will read incorrectly if is it improperly
immersed in the bath. “Total immersion” kinematic viscosity thermom-
eters should be used with the bulb and only the mercury column beneath
the surface of the liquid, but with the emergent stem above the surface at
ambient temperatures.

NOTE

Different thermometers have different immersion requirements. Refer to
the information included with the thermometer in use for specific instruc-
tions.

Figure 12: Installing the motor

Summary of Contents for CT-1000

Page 1: ...ting viscometer tubes thermometers 9 Filling the bath 10 Draining the Bath 11 3 BATH OPERATION 13 Turning on the CT 1000 13 Setting the bath temperature 14 Fixed temperature adjustments 14 Variable te...

Page 2: ...PA USA F APPENDIX F THERMOMETRY 27 KINEMATIC VISCOSITY AND TEMPERATURE 27 ASTM THERMOMETER TABLES 28 ASTM D 445 CHECKING THE ICE POINT 29 NBS MONOGRAPH 150 REPAIRING SEPARATED MERCURY COLUMNS 31 G WAR...

Page 3: ...bility and ease of use it is also suitable for any other application where temperatures must be maintained within close tolerances Temperature range The CANNON CT 1000 will maintain temperatures of 20...

Page 4: ...ps provide glare free illumination of the bath A solid state control circuit equipped with a stainless steel encased thermistor provides proportional temperature control Amotor driven stirrer ensures...

Page 5: ...ns who use the CT 1000 should follow these basic safety procedures Do not place the CT 1000 system on an unstable cart or stand The CT 1000 should be placed on a stable laboratory table or bench If an...

Page 6: ...4 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 7: ...st screwdriver Allen wrench seven hole covers rubber thermometer holder and instruction manual Box 2 contains the bath housing including the electronics drawer Box 2 also contains a smaller box with t...

Page 8: ...ove the front top cover and rear top cover see Figure 4 and 5 Use caution when removing the rear top cover because the tempera ture control probes and heating elements are attached to it Glass jar ins...

Page 9: ...binet 5 Lower the glass jar into the cabinet so it seats evenly on the rubber support ring see Figure 8 6 Unwrap the two glass panels Place the thinner of the two piece of glass in the slot closest to...

Page 10: ...with the bath until the top of the jar forms a tight seal with the covers see Figure 10 Make sure you tighten the set screws uniformly so the jar remains level 15 Run the AC cord through the rear pane...

Page 11: ...before filling the bath with fluid Inserting viscometer tubes thermometers The top cover of the CT 1000 contains seven apertures 51 mm 2 in diameter for the insertion of viscometer tube holders Two a...

Page 12: ...to add fluid until the bath liquid level has risen to approxi mately 40 mm 1 5 of the top of the jar 4 Turn the instrument power ON and incrementally heat the bath to desired control temperature while...

Page 13: ...le container to hold all of the liquid drained from the bath approximately 20 22 liters or 4 5 5 gallons Make sure that the bath liquid is within 10 C of ambient temperature Then insert a tube into th...

Page 14: ...12 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 15: ...ters to match their typeface on the panel Turning on the CT 1000 1 Turn the Power BATH switch on Figure 14 A NOTE The bath heaters will not be activated if there is no liquid in the bath Figure 14 Bat...

Page 16: ...will glow continuously until the bath temperature is approximately that selected with the Temperature SELECT switch when the light will begin to blink The light above the Temperature Adjust control w...

Page 17: ...nd allow several minutes for the bath to equilibrate NOTE If a desired bath temperature is within two or three degrees of one on the Temperature SELECT dial it is sometimes possible to attain the tem...

Page 18: ...16 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 19: ...Check for working motor by connecting to outlet with the proper voltage see motor label for correct voltage Check connections for sensors on rear panel Check temperature setting it must be above exis...

Page 20: ...18 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 21: ...r shaft run out and impeller blade alignment by placing the motor on a soft surface with impeller facing up Connect the motor to the appropriate AC power source and observe the impeller and shaft If r...

Page 22: ...ame side of a shaft as the first mark and is about the same length the bending was not great enough to alter the condition and should be repeated with more force applied If the mark has shifted to the...

Page 23: ...rovides a list of operating temperature ranges and some liquids suitable for use in these ranges C 0 0 2 o t C 0 0 1 S N O I T P O D I U L F H T A B e r u t a r e p m e T C e g n a R s d i u q i L h t...

Page 24: ...ed Because in some cases there is a tendency for algae formation some degree of water treatment may be necessary Water can be used at temperatures close to the boiling point but water replenishment to...

Page 25: ...other than one that is available on the Temperature SELECT switch is to be accessed frequently it may be possible to obtain this temperature by adjusting the trimpot that coincides with the Tempera t...

Page 26: ...24 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 27: ...10 LAMINATEDSAFETYGLASS P27 1320 PLATEGLASS P27 2190 LIGHTEDSWITCHROCKER120V P27 2220 PANELMOUNTLED P27 2230 TEMPERATURERESETSWITCH P27 2230 11 LAMPRESETSWITCH P27 2250 SOLIDSTATERELAY25A P27 2280 BAL...

Page 28: ...26 CANNON CT 1000 Constant Temperature Bath Revision 3 1a February 1998 CANNON Instrument Company P O Box 16 State College PA USA This page intentionally left blank...

Page 29: ...ter CANNON Instru ment Company recommends the use of a calibrated ASTM kinematic viscosity thermometer ASTM Thermometers Each ASTM kinematic viscosity thermometer measures only 3 degrees on a scale su...

Page 30: ...vailable from CANNON Instrument Company ASTM CENTIGRADE THERMOMETERS TYPE CATALOGUE RANGE 74C 9311 K47 55 4 to 52 6 C 73C 9311 K45 41 4 to 38 6 C 126C 9311 K77 27 4 to 24 6 C 127C 9311 K81 21 4 to 18...

Page 31: ...83 5 to 88 5 F 28F 9311 L10 97 5 to 102 5 F 46F 9311 L45 119 5 to 124 5 F 29F 9311 L17 127 5 to 132 5 F 47F 9311 L52 137 5 to 142 5 F 48F 9311 L59 177 5 to 182 5 F 129F 9311 L99 197 5 to 202 5 F 30F 9...

Page 32: ...ewar vessel with the crushed ice and add enough distilled and preferably precooled water to form a slush Do not float the ice Place the thermometer into the slush packing the ice gently around the ste...

Page 33: ...e of centrifu gal force usually serves to unite the mercury in the bulb Care must be taken to warm the top of the bulb first so pressures in the bulb due to expanding mercury may be relieved B If ther...

Page 34: ...of the bulb by rotating the thermometer and tapping it against the palm of the hand When the entire surface has been washed rotate the bubble to the top of the bulb and reunite the mercury as describe...

Page 35: ...damages or any damage to plant personnel equipment or products directly or indirectly resulting from the use or misuse of any product sold by the Company except as set forth in and limited by the fore...

Page 36: ...2139 High Tech Road State College PA 16803 USA Please include the following Required information The Return Authorization number RA The name and telephone number of the person at your company to conta...

Reviews: