29
CANNON
®
CT-1000 Constant Temperature Bath
Revision 3.1a—February, 1998;
CANNON
®
Instrument Company
P.O . Box 16 • State College, PA • USA
ASTM FAHRENHEIT THERMOMETERS
Type
Catalogue #
Range
74F
9311-L80
-67.5 to -62.5°F
73F
9311-L73
-42.5 to -37.5°F
126F
9311-L98
-17.5 to -12.5°F
72F
9311-L66
-2.5 to +2.5°F
128F
9311-L97
29.5 to 34.5°F
44F
9311-L31
66.5 to 71.5°F
45F
9311-L38
74.5 to 79.5°F
118F
9311-L94
83.5 to 88.5°F
28F
9311-L10
97.5 to 102.5°F
46F
9311-L45
119.5 to 124.5°F
29F
9311-L17
127.5 to 132.5°F
47F
9311-L52
137.5 to 142.5°F
48F
9311-L59
177.5 to 182.5°F
129F
9311-L99
197.5 to 202.5°F
30F
9311-L24
207.5 to 212.5°F
110F
9311-L87
272.5 to 277.5°F
NOTE
International shipments may be subject to special shipping regulations.
ASTM D 445 — checking the ice point
Frequency
To achieve an accuracy of ± 0.02°C for calibrated kinematic viscosity
thermometers, a check at the ice point must be made. New thermometers
should be checked monthly for the first six months, then once every six
months