7-138
COPYRIGHT © 2001 CANON INC. CANON CLC1000/1000S/3100 REV.2 MAY 2001 PRINTED IN JAPAN (IMPRIME AU JAPON)
7. TROUBLESHOOTING IMAGE FAULTS/MALFUNCTIONS
49
The cassette heater fails to operate.
Cause
Environment
sensor
Thermal switch
(THS4, THS6),
Cassette heater
(H4, H6)
Connector,
Wiring
Step
1
2
3
Checks
Is there a significant discrepancy
between the temperature/humidity
indicated under ‘DISP’ > ‘ANALOG’
in service mode and the readings of
a thermometer/hygrometer in the
room?
Turn OFF the power switch, and
disconnect the connector J6. Set the
meter to the
×
1
Ω
range. Does the
index of the meter swing when the
probe of the meter is connected to
the terminals of the following con-
nectors?
Upper Cassette Heater
J6M1-1 and J6M1-3
Lower Cassette Heater
J6M2-1 and J6M2-3
Are the connection and the wiring
between the following connectors
normal?
J61, J62, power-saving switch, J6,
J37
Set the meter to the 12VDC range.
Yes/No
NO
NO
NO
YES
Action
Replace the sensor.
Replace the cassette
heater.
Re-connect them.
Check the wiring from the
relay RL1 to the power
saving switch and SSR2;
if normal, replace SSR2.
Summary of Contents for 1000S
Page 12: ......
Page 30: ......
Page 44: ......
Page 86: ......
Page 254: ......
Page 372: ......
Page 374: ......
Page 418: ......
Page 438: ......
Page 442: ......
Page 754: ......
Page 764: ......
Page 766: ......
Page 840: ...0501GR PRINTED IN JAPAN IMPRIME AU JAPON This publication is printed on 100 reprocessed paper ...