5. Basic Photography
-12-
5.1.2 Move the Stage Unit to Get Started
Press the stage unit lock button to release the lock.
Grasp the operation lever, and pull the stage all the way to the
right front, and then press the stage lock button to lock it in
place.
5.1.3 Adjust the Viewfinder
Note: Be sure to adjust the diopter of the viewfinder so it matches the diopter of photographer’s
eye before taking photographs in the mydriatic mode. Otherwise, focusing will not be able
to be performed correctly.
1
Turn the viewfinder diopter compensation ring
counterclockwise (to the positive end) until it
stops.
2
Look into the viewfinder with one eye.
3
Gradually turn the diopter compensation ring
clockwise (toward negative direction) until the
crosshairs in the field of view appear sharpest.