7. Daily Inspection and Maintenance
-40-
7.1.2 Turning ON the Power
Turn on the power of all the equipment which has been connected as a system.
Dig
ita
l came
ra
Check that the digital camera
is firmly attached and the
cables are connected
correctly.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there is any
problem.
Lo
cation
Inspection
Result
Remedy
Month
Day
Month
Day
Month
Day
Lo
ca
tio
n
Inspection
Result
Remedy
Month
Day
Month
Day
Month
Day
Main unit
Check whether the
POWER
lamp lights.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if the POWER lamp
fails to light even though the
power cable is connected
properly.
Check that there is no dirt or
scratches on the objective
lens.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Clean the lens if it is dirty
(see
page 44)
.
Contact a Canon representative or
distributor if it is scratched.
Place a sheet of paper in front
of the objective lens, turn the
brightness adjuster, and check
that the brightness for
observation changes (on the
retinal observation screen).
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are no
changes.
Place a sheet of paper in front
of the objective lens and press
the shutter release button to
check that the flash is emitted.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if no flash is emitted.
Set to fluorescein
angiography mode and press
the
TIMER/C
switch to check
that the timer starts and stops
properly.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are any
problems.
Check that the chin rest
moves smoothly up and down
when the
CHIN REST
switch
is pressed.
Pass/
Fail
Pass/
Fail
Pass/
Fail
Contact a Canon representative or
distributor if there are any
problems.