2. Installation Manual
- 41 -
(8) Table Setup Settings
1) Purpose
Adjust the CXDI operation unit's TABLE SETUP to match the exposure conditions (X-
ray tube voltage, X-ray tube current, msec or mAs value) of the X-ray generator.
2) Procedure
2-1) Start the CXDI system.
2-2) Open the TABLE SETUP Change window from the Normal Exposure window.
System
⇒
SETUP MENU
⇒
SYS. SETUP
⇒
TABLE SETUP
2-3) Select the tabs to be changed and change the X-ray tube voltage, X-ray tube current,
and msec value data to match the exposure conditions of the X-ray generator.
* See the operation manual for the details of settings.
2-4) After finishing the changes, return to the Normal Exposure window and check that the
TABLE SETUP has been changed.