imageFORMULA DR-M160II Service Guide
imageFORMULA DR-M160II Service Guide
June 2014
Page 6
IV.
Servicing Notes
A.
Power and Plug Requirements
Canon USA strongly suggests that dedicated and properly grounded outlets
be provided for the imageFORMULA DR-M160II scanner.
The imageFORMULA DR-M160II requires a 120V, 60 Hz NEMA 5-15R
receptacle for proper operation. Before installation, confirm that the proper
dedicated NEMA receptacles are available for these devices.
Table 7 – Power and plug Requirements
imageFORMULA DR-M160II
NEMA 5-15R
IMPORTANT
We recommend an additional standard 120 V/15 A outlet for service
tools, such as a laptop computer or vacuum that may be used when
servicing or configuring the machine.
Use only dedicated and properly grounded outlets for the Scanner. Do
not use extension cords. The ground connection serves to provide the
internal electronics with a reference voltage. Faulty or poor ground
sources will cause this reference voltage to fall into a range that no
longer serves as a reliable reference voltage. The internal logic and
programming of the imageFORMULA Scanner will not perform reliably
because there is an insufficient difference between the internal
operating signal voltages and the poor ground reference signal. A
qualified electrician can measure and provide the ground source that
the imageFORMULA Scanner machines or any computer controlled
office equipment requires.
Before installation, confirm that all necessary receptacles are available.
Phase converters and step down transformers are not supported.
Table 8 – Power Requirements for the Main unit and Accessories
Part or Accessory
Power Supply
Power Receptacle
imageFORMULA DR-M160II
120 V/15 A outlet
NEMA 5-15R
imageFORMULA Flatbed Scanner Unit 101
imageFORMULA Flatbed Scanner Unit 201