The VistaCool™ Direct-to-Drain Cooling System for Autoclave Wastewater is a patent-pending system designed 
to eliminate the need for autoclave condenser/ waste bottles and condensation tanks by cooling the exhaust 
water and automatically sending it directly down the drain.

Introduction

IMPORTANT INFORMATION

Please read this entire manual before proceeding with installation and operation, and always follow 
local plumbing codes. This system has been verified for installation on SciCan STAT

IM

®

 2000 and 

5000 Autoclave models only.

S6155 

Inline Shut-Off Valve

S6135 

Angle Stop Adapter

S7508 

Inline Thermal  

Sensor Assembly

VC250-A 

VistaCheck™ Dual Check 

Backflow Preventer

S7539 

Cabinet Grommet

x4

S6119 

3/8" x 3/8" x 1/4" Union Tee

S9125 

Wall-Mount C-Clip (3.5”)

x2

TU902 

Cold Water Feed Line 

1/4" OD x 1/8" ID 

Blue Transparent Polyurethane

12

'

TU944 

Condensate Drain Line 

1/4" OD x 0.170" ID 

Black Opaque LLDPE

12

'

TU942 

Coolant Overflow Drain Line 

1/4" OD x 0.170" ID 

Blue Opaque LLDPE

12

'

TU748 

Teflon™ High-Temperature 

Tubing Feed Line  
1/4" OD x 0.185" ID 

9

'

S6112 

Swivel Plug-In Elbow

x2

S1056 

High-Temperature 

Swivel Plug-In Elbow

S7507 

Drain Line Assembly

Summary of Contents for Crosstex VistaCool V7601

Page 1: ...e used with SciCan STATIM 2000 Autoclave SciCan STATIM 5000 Autoclave Includes CSA Certified VistaCheck Dual Check Backflow Preventer PATENT PENDING Direct to Drain Cooling System for Autoclave Wastew...

Page 2: ...ation on SciCan STATIM 2000 and 5000 Autoclave models only S6155 Inline Shut Off Valve S6135 Angle Stop Adapter S7508 Inline Thermal Sensor Assembly VC250 A VistaCheck Dual Check Backflow Preventer S7...

Page 3: ...119 or S6135 Cold Water Line Fitting 1 2 3 4 9 S7503 Thermal Regulator Assembly 10 4 5 S6155 Inline Shut Off Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU942 Coolant Overflow Drain Line 7 8 TU944...

Page 4: ...er 3 to 4 piece of blue poly tubing and connect one end to the outlet end of the inline valve and the other end to the inlet of the supplied CSA certified VistaCheck Dual Check Valve Backflow Prevente...

Page 5: ...ets have been included Simply drill a 7 8 pilot hole through the metal or wood partition then push the smooth plastic grommet into the holes The hole entering the cabinet where the system is installed...

Page 6: ...abinet partition holes If necessary use the high temperature swivel elbow included with the system S1056 to prevent the high temp line connected to the STATIM Autoclave from kinking Connect the black...

Page 7: ...osition Check for blockages in the tubing Autoclave is displaying a cycle fault Exhaust port line is blocked Check for kinks in the tubing from the autoclave to the system manifold Check that water is...

Page 8: ...if the unit is used for purposes other than for what it was designed and manufactured RETURN OF REPLACED COMPONENTS Any item to be replaced under this Warranty must be returned to Crosstex in Phoenix...

Reviews: