5

Cabinetry

The VistaCool™ Direct-to-Drain Cooling System for Autoclave Wastewater is typically installed in the base/
sink cabinet of sterilization centers or at the sterilization tower if plumbed with separate water and drain 
connections. The 1/4" O.D. Teflon™ High-Temperature Tubing will need to run from the output port of 
the STAT

IM

®

 Autoclave directly to the system. This exhaust line should run in a progressively descending 

trajectory, per STAT

IM

 Autoclave requirements.

In many cases, holes will need to be drilled through one or more partitions in order to get the tubing 
from the autoclave to the VistaCool Direct-to-Drain Cooling System for Autoclave Wastewater. In order 
to protect the tubing from kinking or damage, grommets have been included. Simply drill a 7/8" pilot 
hole through the metal or wood partition then push the smooth plastic grommet into the holes. The hole 
entering the cabinet where the system is installed should be slightly above the top of the system so the 
tubing has a natural gravity angle to help prevent pools of liquid in the tubing.

AUTOCLAVE

USE 7/8

"

 

HOLE SAW

Install the system

 

along dashed path

90

º

Acceptible

Positions

If installed horizontally,
coolant overflow port must
be UP in 12 o’clock position

Coolant overflow

always at 12 o’clock

The system may be installed horizontally or vertically. If installed horizontally, the coolant overflow port must be UP in 
the 12 o’clock position. Do not install the system upside down or at an angle other than shown here as acceptable. Affix 
both included C-clips securely to the desired mounting surface using appropriate screws to ensure the system does not 
move during operation.

Summary of Contents for Crosstex VistaCool V7601

Page 1: ...e used with SciCan STATIM 2000 Autoclave SciCan STATIM 5000 Autoclave Includes CSA Certified VistaCheck Dual Check Backflow Preventer PATENT PENDING Direct to Drain Cooling System for Autoclave Wastew...

Page 2: ...ation on SciCan STATIM 2000 and 5000 Autoclave models only S6155 Inline Shut Off Valve S6135 Angle Stop Adapter S7508 Inline Thermal Sensor Assembly VC250 A VistaCheck Dual Check Backflow Preventer S7...

Page 3: ...119 or S6135 Cold Water Line Fitting 1 2 3 4 9 S7503 Thermal Regulator Assembly 10 4 5 S6155 Inline Shut Off Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU942 Coolant Overflow Drain Line 7 8 TU944...

Page 4: ...er 3 to 4 piece of blue poly tubing and connect one end to the outlet end of the inline valve and the other end to the inlet of the supplied CSA certified VistaCheck Dual Check Valve Backflow Prevente...

Page 5: ...ets have been included Simply drill a 7 8 pilot hole through the metal or wood partition then push the smooth plastic grommet into the holes The hole entering the cabinet where the system is installed...

Page 6: ...abinet partition holes If necessary use the high temperature swivel elbow included with the system S1056 to prevent the high temp line connected to the STATIM Autoclave from kinking Connect the black...

Page 7: ...osition Check for blockages in the tubing Autoclave is displaying a cycle fault Exhaust port line is blocked Check for kinks in the tubing from the autoclave to the system manifold Check that water is...

Page 8: ...if the unit is used for purposes other than for what it was designed and manufactured RETURN OF REPLACED COMPONENTS Any item to be replaced under this Warranty must be returned to Crosstex in Phoenix...

Reviews: