background image

7

Starting & Using the System

Turn on the main water supply and check for leaks.

Slowly open the inline water supply valve fully and check for leaks.

Prime the system by inserting a small screwdriver in the viewport hole on 
the brass stem of the thermal regulator and pushing the piston away from 
the system body to activate cold water flow. The system is primed when 
water can be heard flowing through the 1/4" solid blue overflow tubing 
and running to the drain.

Ensure the system is installed in an elevated position above the drain connection 
as described previously to ensure proper flow of the condensate line, as required 
by STAT

IM

®

 Autoclave instructions. 

The coolant overflow fitting should always be 

UP in the 12 o’clock position if the system is installed horizontally

.

Run three consecutive autoclave cycles to ensure everything is working properly.

1

2

3

4

5

Troubleshooting

System is not draining

•  Ensure water supply valve is in the ON position.

•  Check for blockages in the tubing.

Autoclave is displaying a 
cycle fault

•  Exhaust port line is blocked. Check for kinks in the tubing from the autoclave 

to the system manifold.

•  Check that water is running from the tank coolant overflow line to the drain 

adapter.

•  Make certain the water supply valve is on. If there is no water to cool the 

coil in the tank, the inline thermal sensor located in the condensate line will 
respond by closing the pathway to alert you of the problem and protect the 
plumbing from steam. 

The black condensate line 
appears to be blocked

•  The inline thermal sensor has been activated, blocking condensate flow to 

the drain, caused by a lack of coolant water in the tank that overheated the 
condensate. Overheated condensate will cause the wax motor to expand, 
closing the outlet. To reset the inline sensor and enable flow to resume, 
allow the system to rest and cool naturally. The inline sensor’s wax-motor 
piston will retract, opening the outlet port. 

Water leaking under counter

•  Check all tubing and fitting connections

•  Push in on all push-to-connect fittings to ensure the tubing is fully engaged 

in the fitting “O” ring.

Summary of Contents for Crosstex VistaCool V7601

Page 1: ...e used with SciCan STATIM 2000 Autoclave SciCan STATIM 5000 Autoclave Includes CSA Certified VistaCheck Dual Check Backflow Preventer PATENT PENDING Direct to Drain Cooling System for Autoclave Wastew...

Page 2: ...ation on SciCan STATIM 2000 and 5000 Autoclave models only S6155 Inline Shut Off Valve S6135 Angle Stop Adapter S7508 Inline Thermal Sensor Assembly VC250 A VistaCheck Dual Check Backflow Preventer S7...

Page 3: ...119 or S6135 Cold Water Line Fitting 1 2 3 4 9 S7503 Thermal Regulator Assembly 10 4 5 S6155 Inline Shut Off Valve 5 VC250 A VistaCheck Backflow Preventer 6 TU942 Coolant Overflow Drain Line 7 8 TU944...

Page 4: ...er 3 to 4 piece of blue poly tubing and connect one end to the outlet end of the inline valve and the other end to the inlet of the supplied CSA certified VistaCheck Dual Check Valve Backflow Prevente...

Page 5: ...ets have been included Simply drill a 7 8 pilot hole through the metal or wood partition then push the smooth plastic grommet into the holes The hole entering the cabinet where the system is installed...

Page 6: ...abinet partition holes If necessary use the high temperature swivel elbow included with the system S1056 to prevent the high temp line connected to the STATIM Autoclave from kinking Connect the black...

Page 7: ...osition Check for blockages in the tubing Autoclave is displaying a cycle fault Exhaust port line is blocked Check for kinks in the tubing from the autoclave to the system manifold Check that water is...

Page 8: ...if the unit is used for purposes other than for what it was designed and manufactured RETURN OF REPLACED COMPONENTS Any item to be replaced under this Warranty must be returned to Crosstex in Phoenix...

Reviews: