The following COMMAND table describes the available commands:
WIRELESS MODE COMMAND table
Flash count
Command
Description
1 flash
STATUS
Shows the device STATUS. As answer the led perform a flash sequence
as reported in the STATUS table. If the device is performing the TEST
(
refer to TEST command
) this command stops it.
2 flashes
TEST
Enter in TEST mode and transmits status and measurements every 5
seconds. If the device is in STANDBY mode or it is out of radio range,
this command forces the connection procedure to the WSN and the
return to the operative mode. The TEST stops after 120 seconds.
During TEST, the led continuously shows the STATUS to monitor the
received radio signal quality.
CAUTION: Measures acquired during TEST phase are NOT saved.
3 flashes
ENROL
Association to the network: must be used when the device has not yet
been included in a network, starts the entry and association procedure
to the gateway (
refer to
WineCap System - User Manual R30
).
4 flashes
+ 4 flashes
STANDBY
Temporary device deactivation: the device is stopped. The sampling
process and the radio are/is. turned off losing the connection to the
network. To reactivate, a TEST command is necessary. The STANDBY
command must be given twice to confirm it: at the first sequence the
led flashes alternating RED and GREEN lights, waiting for the second
confirm sequence within 15 seconds. At the command execution the
led flashes as the STANDBY status (
refer to
).
5 flashes
+ 5 flashes
FACTORY RESET
The device performs the memory deleting procedure and goes in
STOP status. All samples, configuration and wireless network data
associated are LOST. To reactivate the device a new association and
configuration procedure is necessary (
ENROL command
). Also in this
case, the FACTORY RESET command must be given twice to confirm
it. At the command execution the led flashes as the
“PROBE/DATALOGGER NOT ASSOCIATED” status
(
refer to “
).
6 flashes
+ 2 flashes
CH2 AND CH3
ZEROING
The device performs the inclination channels X and Y axis zeroing
procedure,
fixing the device’s absolute position and the consequent
AXIS ROTATION aligning the internal references and saving the ZERO
position.
6 flashes
+ 6 flashes
ZEROING
CANCELLATION
The device performs the cancellation procedure of zeroing operations
on channels 2 and 3. All the initial values are restored.
Picture 5 - Command table