10.0 Specifications
10.0
Specifications
Carbolite Gero reserves the right to change the specification without notice.
All models have cylindrical elements with wire mounted in the surface of the insulation
material.
All models can accept work tubes up to a maximum outside diameter of 60 mm.
All models have a maximum operation temperature of 1200°C (1100°C continuous).
Model
Max Temp
(°C)
Max Power
(kW)
Minimum
Work
Tube
Length
(mm)
Tube length
for use with
modified
atmosphere
(mm)
Heated
Length
(mm)
Weight
(kg)
Vertical 3-zone tube furnaces.
EVZ 12/450B
1200
2.0
600
900
450
31
Work tube adaptors, extended work tube supports, insulation plugs and gas/ vacuum
end seals are available from Carbolite Gero for work tubes with outside diameters of 32
mm, 46 mm and 60 mm.
10.1
Environment
The furnaces contain electrical parts and should be stored and used in indoor conditions
as follows:
Ambient temperature working range
Temperature:
5 °C to 40 °C
Note: when operating the furnace at temperatures close to the maximum and the
ambient temperature is above 30 °C, the allowed external temperature defined in EN
61010-1:2010 may be exceeded.
Relative humidity
Maximum 80% up to 31 °C, decreasing linearly to 50% at 40 °C.
Important safety notice:
After transportation or storage in humid conditions, the furnace could fail
to meet all the safety requirements of BSEN 61010-2-010 until it has
completed the drying out process to restore its normal condition.
Warning:
It cannot be assumed that the furnace will meet all the safety requirements of BSEN
61010-2-010 during the drying out process.