6.3
Electrical Door Interlock (if fitted)
Note:
To open the door, the electrical supply to the product must be switched on!
The electrical door interlock is situated on the
underside of the retort door. It is linked to a relay
within the main temperature controller and can be
configured to operate based on a temperature
limit.
Note:
If a gas system is fitted, the electrical door interlock will also activate if the
gas is flowing.
The door interlock is activated when the temperature of the product is above a pre-set
level, usually 80°C.
When the product is above the pre-set temperature, the indicator lamp will illuminate
and the operator will be unable to open the door.
The door lock will automatically deactivate when the product drops below the pre-set
temperature level.